top of page
DSC Type:
Measurement Principle:
Cooling Accessory:
Atmosphere:
Calorimetric Performance:
Single furnace
Heat flux
Intracooler (PerkinElmer)
Nitrogen and argon are available for sample purge
Dynamic range: ±175 mW
Accuracy: ± 2%
Precision: ± 0.1%
Digital resolution: 0.02 µW
Temperature Performance:
Range: -80 to 445 °C
Accuracy: ± 0.1 °C
Precision: ± 0.02 °C
Data points/sec: 10
Controlled heating rates: 0.1 to 100 °C/min
Controlled cooling rates: 0.1 to 100 °C/min
Intracooler cooling times (100 °C to 0 °C): < 4 min, for fast sample turnaround
Advanced Mode:
Software:
Modulated Temperature DSC (MT-DSC)
Pyris software (version 13) with Isothermal Kinetics, Scanning Kinetics, Specific Heat and Purity software packages
bottom of page